Measuring instruments of parameters of semiconductor devices IPPP-3, IPPP-3/1

Печать

Описание

  • Measurement and analysis of current-voltage characteristics (CVC) in the ranges of currents up to 20 A and voltages up to 2000 V
  • Dynamically displaying the current-voltage characteristic graph when manually adjusting the voltage of the repeated scan
  • A single measurement of the current-voltage characteristic by an internal or external trigger signal
  • The meter contains channels C (collector, drain), B (base, gate), E (emitter, source), S (substrate)
  • Types of sweep signals in channel С:
    • half-wave sinusoidal resolution (10 ms)
    • three periods of sinusoidal voltage (60 ms)
    • pulses of increasing amplitude (0.2 ms ÷ 2 ms)
    • constant pressure
  • The basic measurement error in channels C, B, S - 1.5%
  • Taking measurements does not require programming knowledge
  • Measurement of parameters of semiconductor devices (including on a wafer), defect analysis
  • Limitation of current and voltage at the electrodes of the test object according to predefined thresholds in channels C, B, S
  • Display of measured and calculated characteristics in the form of graphs and tables
  • Saving tables electronically
  • Formation of a report on the results of measurement, calculation and permissible control of parameters with an indication of suitability
  • The presence of a library of standard tests for measuring the parameters of standard semiconductor devices
  • Ability to create a library of user tests by setting parameters for measurement conditions
  • The ability to create your own measurement models according to the principle: product, test item, test
  • Creating your own archive of reference graphs with the ability to impose a reference I – V characteristic on a measured characteristic
  • Marker measurements, tangent plotting
  • Inverted polarity of the X and Y axes
  • Loop Compensation, Screen Magnifier Mode

Specifications

  • Channel C (collector, drain)

Voltage Source (Uc):

Collector voltage

 

        Maximum

       (peak) current

Duration
impulses

R

Int.

 

IPPP-3

IPPP-3/1

0 - +5 V

--

20 A

0.2 mс

»0.07 Ohms

0 - -5 V

--

20 A

0.2 mс

   »0.07 Ohms

0 - ±20 V

10 A

10 A

0.2 mс

»0.4 Ohms

0 - ±100 V

1,5 A

2 A

0.5 mс

»2.0 Ohms

0 - ±500 V

0,25 A

0,4 А

1 mс

60 Ohms

0 - -±2000 V

0,06 A

0,06 A

2 mс

4800 Ohms

Current Meter (Ic):

- ranges from ± 10 nA to ± 20 A (± 50 A IPPP-3/1), according to series 1, 2, 5, 10;

- minimum sampling - 0.005 nA;

- measurement error:

                   ± [(1.0¸1.5)% of I + 0.5% of I], with Ic ≤ 2A;

                   ± [9% of Ii + 1% of Ip], at Ic> 2A.

 

Voltage Meter (Uc):

- ranges from ± 0.5 V to ± 2000 V, according to series 1, 2, 5, 10;

- minimum discretion - 0.2 mV;

- measurement error: ± (1% of Uи + 0.5% of Uп),

where Uп (Iп) is the upper limit of the voltage (current) range;

Uи (Iи) - voltage (current), measured at the load;

Uf (If) - formed voltage (current).

  •  Channel B (base, gate)

Type of output signal current (voltage):

- stepwise variable signal with offset;

- impulse steps with impulse bias;

- constant level.

Current (voltage) stage generator:

- number of steps - from 0 to 10; 
- current steps (Istep) from 2 nA / stup to 0.5 A / stup;
- voltage steps (Ustep) from 50 mV / stup to 2 V / stup, according to a number 1, 2, 5, 10;
- tenfold reduction of a step (x0.1);
- bias current (voltage) range - ± 10 · Istep (± 10 · Ustep);
- maximum constant output current Iout - ± 2.5 A, with Uout ≤ 4V;
- maximum output voltage - ± 40 V, at Iout £ 0.1 A;
- output resistance of at least 1010 Ohms, with Istep = 2 nA / stup.

Current Meter (Ib):

-ranges from ± 40 nA to ± 10 A, according to series 1, 2, 4, 10;

- minimum discretion - 0.02 nA;

- error in the formation (measurement) of current -

± [(1.0¸1.5)% of Iph (Ii) + 0.5% of Ip], at Ib £ 2.5 A.

Voltage Meter (Ub):

- ranges from ± 1 V to ± 40 V, according to series 1, 2, 4, 10;

- minimum discretion - 0.5 mV;

- the error in the formation (measurement) of voltage -

± (1% of Uf (Ui) + 0.5% of Uп).

 

  • Channel E (emitter, source) - zero potential (common).

 

  • Channel S (substrate)

Maximum output current (Iout) - ± 10mA, with Uout £ 40V;

For other parameters see “Channel B”.

 

Contents of delivery:

IPPP-3 or IPPP-3/1 measuring instrument.

The portable device for installation of the test object, equipped with a transparent cover and a removable metal screen.

Two-wire connector for connecting two- and three-terminal devices, two-wire contact clamps (3 pcs.), Single-wire probe clamps (4 pcs.) For connecting to the electrodes of the test object.

Cable connector for connecting to probes (two-wire connection).

 

General characteristics:

12.1 "LCD screen with a resolution of 1024 x 768, keyboard, mouse.

USB, Ethernet.

Supply voltage: 230 V / 50 Hz., Power: 450 V · And.

Operating temperature range: from + 5 ° to + 40 ° С.

Dimensions (W x H x D): 470 x 330 x 570 mm, weight: 37 kg